RT100-B

RT100 electronics application

The RT100-B LIBS system is an ideal tool for material discrimination. It houses an ICCD-coupled broadband spectrometer that allows users to collect LIBS spectra from 200 nm to 900 nm with a single laser shot. By coupling broadband LIBS spectra with ASI's proprietary chemometrics software package, the RT100-B can discern even the most complex of materials with a high degree of accuracy.

The RT100-B’s wide variety of options makes it configurable to work well both in laboratory and production environments. Based on the RT100-HP's robust, proven platform, the RT100-B continues to set the industry standard for accurate and repeatable broadband LIBS analysis.

RT100-B model LIBS system

An integrated laser ablation system

The RT100-B LIBS system uses an Nd:YAG laser for ablating samples. The laser wavelength is carefully chosen to provide the best analytical results for each customer’s unique application.

Optimized plasma light collection

The RT100-B incorporates a unique design that efficiently collects light emitted from the LIBS plasma. This capability is crucial for achieving highly accurate and repeatable test results.

Designed for flexible sampling protocols

The RT100-B houses a 3D (X- Y- Z) translational stage with excellent scan ranges. Various spatial sampling patterns can be programmed to yield a highly automated LIBS measurement experience.

ASI hardware technology enhances data precision

One of the critical system requirements for achieving improved LIBS data precision is a consistent laser focus to the sample surface. With patent-pending ASI hardware technology, our customers are assured that their LIBS measurements are accurate and repeatable.

Echelle Spectrometer with ICCD detector

The RT100-B incorporates an Echelle spectrometer with ICCD camera to reduce background interference and enhance detection sensitivity. Independent ICCD camera control includes detection parameter settings, such as gate delay, gate width and signal gain, which allow the operator to achieve superior signal-to-noise and signal-to-continuum performance. Using an Eschelle spectrometer allows the RT100-B to capture LIBS spectra regions ranging from UV to NIR with a single laser shot. ASI’s technical sales staff helps customers choose the right ICCD camera for their specific Echelle-based LIBS analyses.

ASI chemometric software package

ASI’s proprietary chemometric software utilizes a set of highly sophisticated statistical algorithms to add material discrimination capability to our LIBS instruments. Proven through extensive testing of a wide range of materials with different interferents, ASI’s chemometric software package adds another level of analytical capability for our customers.

Calibration standards development

LIBS is a versatile analytical technique and, with the proper sampling methodologies, LIBS can accurately quantify many elements accurately down to the ppm level. As with many quantitative analytical techniques, the availability of appropriate calibration standards ensures measurement accuracy. With our scientific team's 80-plus years of combined laser ablation knowledge, ASI is well suited to develop such calibration standards and to recommend the optimal laser and detection parameters for your LIBS analysis. No other LIBS equipment manufacturer can offer this level of technical knowledge and application support.

RT100 LIBS discrimination application

Software control

All our RT100 series systems come with powerful software, Axiom, together with our TruLIBS database. Axiom feature a simple, intuitive graphical interface that allows multiple users to operate the system software with different access privileges. Thus, the same RT100-HP can be used by production line operators to simply run a prepared measurement recipe for routine quality control work, or by the laboratory scientists to perform cutting-edge elemental research.

 

Laser High power Nd:YAG Laser, different wavelength available
optimized for customer application
Imaging 1280x1024 CMOS color camera (minimum); Upgrades available.
Automated X-Y Stage 50 mm x 50 mm travel range, Speed range 0.001-20 mm/sec
 0.1 micron resolution with 0.2 micron repeatability
Automated Z stage 26 mm travel range, Speed range 0.001-20 mm/sec
 0.2 micron resolution with 0.1 micron repeatability
Ablation spot guiding laser Red laser @670 nm
Spectrometer/ Detector Echelle Spectrograph/ICCD system with 200 to 900nm spectral window
Control system PC control running Windows XP and ASI system control software
Particle filter Mounted HEPA air cleaner to purge ablated particles
Physical parameters 29 in/ 736.6 mm (W) x 32 in/ 812.8 mm (D) x 54 in/ 1371.6 mm
with cover lid (H) 350 Lb/ 158 Kg
Laser safety Class I laser product, sample loading area shielded with optical filter.
Laser interlock protected
Site requirements 70°±10°F/ 21°±5°C, 110 (10A)-240 (5A) VAC, 50/60Hz
Options Other hardware options are available for customer-specific applications.
Please contact the factory for available options.